Current location:HOME>PRODUCT>Oil particle counter>Oil Cleanliness Monitor
Oil Cleanliness Monitor

Oil Cleanliness Monitor HM-YZ10

【Introduction】Small size, without pump, small sensor

Product details

The oil cleanliness monitor is a device specifically designed for on-site online measurement of oil contamination levels, using the counting principle of the light blocking method specified by the International Hydraulic Standards Committee. It has the advantages of small size, light weight, fast detection speed, high accuracy, and good repeatability, and can work under harsh conditions such as high temperature and high pressure. Suitable for all transparent solvents such as engine oil, gear oil, transformer oil (i.e. insulation oil), hydraulic oil, lubricating oil, synthetic oil, water-based (water-based hydraulic oil, water ethylene glycol, etc.), alcohols, ketones, etc. It can be widely used in aerospace, petrochemical, transportation and port, steel metallurgy, automobile manufacturing and other fields.

Main Features

Using the principle of photoresist (light blocking) method and high-precision laser sensors, it has small size, high accuracy, and stable performance

Suitable for on-site online detection, capable of real-time monitoring of particle contamination in oil systems

Built in data analysis system, capable of displaying real data of particle size in each channel and automatically determining sample grade

The standard model can directly withstand a pressure of 100 kilograms and can be equipped with a pressure reducing valve for online measurement of ultra-high pressure

Equipped with volume flushing and duration flushing modes, convenient for users to use and maintain the equipment

Built in particle pollution level standards such as ISO4406, NAS1638, SAE4059, GJB420A, GJB420B, Г OCT17216, GB/T14039, etc

Built in calibration function, can calibrate according to standards such as GB/T21540, ISO4402, ISO11171, GB/T18854, etc. One test can provide all built-in standard results

Can independently set all standards and any alarm level to achieve pollution or cleanliness detection

The standard model has RS232, RS485, and USB interfaces, supports standard Modbus protocol, and can be connected to computers, upper computers, printers, PLC systems, or other devices for data monitoring and processing

Ultra large storage, can be stored inside the instrument or on external storage devices

Durable exterior structure, suitable for complex working environments

The bottom in and top out mode is beneficial for minimizing the interference of online bubbles on test results

Continuous testing is possible, and the testing time interval can be set arbitrarily

Dual system of Chinese and English, customers can freely switch, suitable for export

Touch screen or membrane button operation, can be freely switched, and the instrument interface can freely control the switch of the remote printer

Optional connection to industrial control system equipment for remote data monitoring, historical data, and curve query (optional)

Built in moisture and temperature sensor module, capable of simultaneously outputting four types of parameter information (standard model optional)

Technical indicators

Light source: semiconductor laser

Flow rate range: 5-500mL/min

Test sample viscosity: ≤ 650cSt

Online pressure detection: 0.1~10 Mpa (with optional pressure reducing device, the maximum pressure can reach 42 Mpa)

Particle size range: 1-1000 μ m

Standard interfaces: USB interface, RS232 interface, RS485 interface

Data storage: Provides 1000 sets of data storage space and supports USB storage

Sensitivity: 1 μ m or 4 μ m (c)

Limit coincidence error: 40000 particles/ml

Counting volume: 1~999ml

Counting accuracy: ± 0.5 pollution level

Protection level: IP56/IP67

Temperature (optional): Collection range: 1-100 ℃; Measurement accuracy: 1 ℃

Water activity (optional): Collection range: 1-100% RH; Measurement accuracy: 1% RH

Moisture content (optional): Collection range: 1~360ppm; Measurement accuracy: 1ppm

Test interval: 1 second to 24 hours

Test sample temperature: 0~80 ℃

Working temperature: -20~60 ℃

Power supply: AC 220V ± 10%, 50/60Hz or DC9-36V

Weight: 1kg

Volume: 115 × 85 × 60mm